• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER GMBH CMN FEMORAL NAIL, CCD 125, LEFT, 13 MM, 34 CM; ZIMMER NATURAL NAIL SYSTEM CEPHALOMEDULLARY NAILS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER GMBH CMN FEMORAL NAIL, CCD 125, LEFT, 13 MM, 34 CM; ZIMMER NATURAL NAIL SYSTEM CEPHALOMEDULLARY NAILS Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products and therapy date: detail of product: item number: 47248509510, item name: znn, cmn lag screw, 10.5 mm, 95 mm, including set screw, lot #: 2938330.Item number: 47248403750, item name: 5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head, lot #: 63779982.Item number: 47248404050, item name: 5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head, lot #: 63985506.Item number: 47248404250, item name: 5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head, lot #: 63869850.Item number: 47248700205, item name: cephalomedullary nail cap 5 mm height, lot #: 63880389.The manufacturer did not receive x-rays, or other source documents for review.The manufacturer did not receive the device for investigation.The device history records were reviewed and found to be conforming.Additional information has been requested and is currently not available.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available, an updated report will be submitted.Zimmer biomet¿s reference number of this file is (b)(4).
 
Event or Problem Description
It was reported that patient underwent revision surgery due to implant fracture.
 
Additional Manufacturer Narrative
This follow-up report is being filled to relay investigation result.Dhr review: the device manufacturing quality records indicate that the released components met all requirements to perform as intended.Trend analysis: no trend considering the following event is identified: implant fracture.Event description: patient was implanted with a cmn femoral nail on (b)(6) 2018 and underwent revision surgery on (b)(6) 2018 due to fracture.Review of received data: no medical data such as x-rays, surgical notes or any other case-relevant documents received.Device analysis: no product was returned to zimmer biomet for in-depth analysis.Review of product documentation: the compatibility check was performed and showed that the product combination was approved by zimmer biomet.Conclusion: it was reported that the patient was implanted with a cmn femoral nail on (b)(6) 2018 and underwent revision surgery on (b)(6) 2018 due to fracture.The nail was in vivo for approx.5 months.Neither x-rays, operative notes, office visit notes, nor devices or photos of the explanted implants were received; therefore the condition of the components is unknown.It is known that the bone quality of the patient is inferior.The medical history of the patient: aortic aneurysm, cholecystitis and alcoholism.Based on the given information and the results of the investigation, the complaint could not be confirmed as the alleged failure could not be identified or reproduced.The need for corrective measures is not indicated and zimmer gmbh considers this case as closed.Zimmer biomet¿s reference number of this file is (b)(4).
 
Event or Problem Description
Please refer to report 0009613350-2018-01166.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CMN FEMORAL NAIL, CCD 125, LEFT, 13 MM, 34 CM
Common Device Name
ZIMMER NATURAL NAIL SYSTEM CEPHALOMEDULLARY NAILS
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
MDR Report Key8101537
Report Number0009613350-2018-01166
Device Sequence Number11261394
Product Code HSB
UDI-Device Identifier00889024299481
UDI-Public00889024299481
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K091566
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number47-2493-341-13
Device Lot Number2914731
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/05/2018
Supplement Date Received by Manufacturer02/26/2019
Initial Report FDA Received Date11/26/2018
Supplement Report FDA Received Date02/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
REFER TO H10
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age63 YR
-
-