No complaint was received with return of the device.Failure event observed during analysis.Final analysis found the device to be in backup mode.The cause of the device reset was an uncorrectable nmi which could not be reproduced in the laboratory.The device was tested on the bench and using automated testing equipment, and no other anomalies were found.In addition, burn marks were observed on the can.Abbott conducted a review of complaint files under a non-product capa and identified that a report was required to be filed as a result of the subject event.Distributed devices are not affected by the capa.Additionally, no new risks were identified, and no product-related actions are required as a result of this report submission.Furthermore, product performance trending is not impacted by this report, and no new mitigation actions are required as a result of this report.
|