US SURGICAL PUERTO RICO ROTICULATOR; STAPLE, IMPLANTABLE Back to Search Results
Model Number 017614
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/17/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report was originally submitted via e2001015.Evaluation summary: post market vigilance (pmv) led an evaluation of one roticulator 55mm-4.8mm single use stapler.The instrument was received with the pushers advanced and the handle was not locked.Functionally, the instrument was reset and cycled.The jaw properly clamped and unclamped when the approximating lever was operated.The handle locked into place.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all medtronic quality release specifications at the time of manufacture.Replication of the observed condition may occur when the firing stroke is not completed during clinical application.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter: during an open sigmoidectomy, the stapler did not fire.No patient injury.To complete the procedure, a new device was opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTICULATOR
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key8102179
Report Number2647580-2018-05722
Device Sequence Number1599325
Product Code GDW
UDI-Device Identifier10884521047709
UDI-Public10884521047709
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K855047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2018
Device Model Number017614
Device Catalogue Number017614
Device Lot NumberP3G0031X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2018
Initial Report FDA Received Date11/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-