| Model Number |
4002 |
| Medical Device Problem Code |
Solder Joint Fracture (2324)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This monitor is the home version of the smart monitor 2 and the jack on the back of the monitor is for interfacing to a previously sold remote alarm for home use.This jack is not operable with a nurse call system.The user facility has been informed that this interface should not be used with a nurse call system as it is not intended for that purpose.
|
| |
|
Event or Problem Description
|
|
A user facility reported that the nurse call jack on the monitor was not functioning properly.
|
| |
|