| Catalog Number |
389.873 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
11/02/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history records has been requested.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, the patient underwent an anterior lumbar interbody fusion (alif).The fixation pin was used to temporary fix the anterior tension band (atb) plate to the correct position.After having fixed the first three screws, during removal of the fourth fixation pin the tip broke off.Tip was left in place and the final screw was inserted.The final screw was smaller than the remaining screws.Procedure was completed successfully with no delay.Concomitant device reported: screw (part unknown, lot unknown, quantity 3), atb plate (part unknown, lot unknown, quantity 1).This report is for one (1) detachable temporary fixation pin.This is report 1 of 1 for (b)(4).
|
| |
|
Event or Problem Description
|
|
Patient outcome is reported as normal.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If additional narrative: description of event or problem: additional information.A device history record (dhr) review was conducted: part number: 389.873.Synthes lot number: u311794.Supplier lot number: u311794.Release to warehouse date: (b)(6)2018.Manufacturing site: synthes brandywine.Supplier: (b)(4).The raw material was confirmed to be correct per the specification with no (relevant) non-conformance noted.No non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.A product investigation was conducted.Visual inspection:the returned fixation pin was received broken.The fracture occurred within the threads, near the proximal end of the threads.The piece distal to the fracture was not returned.The surface of the fracture displayed no signs of material defect and was irregular with varying angles with respect to the longitudinal axis of the device's shaft.Cosmetic wear consistent with the device's use was observed on the non-threaded shaft; this wear would not affect performance of the device.The received condition agreed with the complaint description.Therefore, the complaint was confirmed.Dimensional inspection: a dimensional inspection of the threads could not be performed due to the received condition of the device.The outer shaft diameter proximal to the break measured and was within specification per relevant drawing.Dhr and material review: review of the dhr revealed no manufacturing issues related to the complaint.A material review revealed no issues related to the material specifications of the device.Document/specification review: relevant drawings for the returned instrument were reviewed:no design issues were observed.Conclusion: the complaint was confirmed with investigation.There were no issues during the manufacture of this product that would contribute to this complaint condition.During the investigation, no product design issues or discrepancies were observed that may have contributed to the complaint condition.The current risk assessment document was found to appropriately address the complaint condition.A root cause could not be determined; however, it is likely that the device experienced excessive and/or off-axis forces.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Therefore, it has been determined that no corrective and/or preventive action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|