| Model Number |
MICL 13.7 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
10/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Word order search: no similar complaint type events were reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.7mm micl13.7, -11.5 diopter, implantable collamer lens into the patients right eye (od) on (b)(6) 2018.On (b)(6) 2018 the lens was removed due to low vault (subluxation).It was reported event was due to patient related factor.There are no plans to implant another icl.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Lens was returned in liquid, in lens vial.Visual inspection found the haptic torn.(b)(4).
|
| |
|