If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device is not being returned, therefore is unavailable for a physical evaluation.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaints of any kind for this lot of devices that were released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the remetrex complaint management system as required under mitek¿s corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
|
It was reported that during an acl recon, a milagro was screwed into to bone.During screwing, the surgeon felt that everything was fine but in the end the screwdriver didn´t work anymore.When he pulled out the screw, it was noted to be partially broken.The larger part stayed in a bone and everything was fine and fixed.The operation was finished and nothing was damaged.No delay in procedure.
|