| Medical Device Problem Codes |
Partial Blockage (1065); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a occlusion (occlusion issue) issue.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long-term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 02-jan-2019 with the following findings: review of the black box data and alarm history reveal multiple records of cs145 occlusion alarms due to high force.On investigation, the pump powered on normally and ez-prime steps successfully completed without any alarms, errors or warnings.Investigation did not duplicate the occlusion complaint.Unrelated to the original complaint, the display screen was noted to be dim/discolored.
|
| |
|