| Model Number |
9733856R |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No patient information provided as no patient was involved in this concern.No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that, while outside of a procedure, the surgeon monitor of the navigation system would blink when a bovie electrical cutter is introduced into the field of navigation.There was no patient present when this issue was identified.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that the navigation system was tested and found to function as designed.
|
| |
|