Device display properly and no missing segment/partial display anomaly noted.However, device had unresponsive buttons due to flattened (creased) act and up buttons dome switch.No unlocked lcd keypad connector noted.Unable to perform operating currents, self-test, a21 error test, rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test or verify prime/fill, rewind anomaly due to unresponsive button.Device had minor scratched lcd window, cracked battery tube threads.(b)(4).
|