MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAL
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Mechanical Problem (1384); Device Difficult to Program or Calibrate (1496); Patient Device Interaction Problem (4001); Priming Problem (4040)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had keypad anomaly and prime anomaly.The customer¿s blood glucose level was unknown.The customer reported that the act button during priming stopped working and had to press multiple times to respond.It was reported that the pump screen had scratches on the display that effected the ability to see screen fully.It was reported that they had to rewind twice during priming of the pump.The customer was advised the pump will need to be replaced.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Device display properly and no missing segment/partial display anomaly noted.However, device had unresponsive buttons due to flattened (creased) act and up buttons dome switch.No unlocked lcd keypad connector noted.Unable to perform operating currents, self-test, a21 error test, rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test or verify prime/fill, rewind anomaly due to unresponsive button.Device had minor scratched lcd window, cracked battery tube threads.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAL
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8103662
Report Number2032227-2018-66409
Device Sequence Number1098320
Product Code OZO
UDI-Device Identifier00643169736085
UDI-Public(01)00643169736085
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-751NAL
Device Catalogue NumberMMT-751NAL
Device Lot NumberA6751NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2019
Initial Date Received by Manufacturer 11/19/2018
Supplement Date Received by Manufacturer11/21/2019
Initial Report FDA Received Date11/26/2018
Supplement Report FDA Received Date11/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age30 YR
Patient Weight70
-
-