No complaint was received with the return of the device.Failure event observed during analysis.Final analysis found the reported inability to communicate with the device was confirmed in the laboratory and was due to electrolyte leakage from the battery.The device was tested on the bench and the electrolyte caused hybrid damage and premature battery depletion.The cause of the leakage is due to a battery anomaly.Abbott conducted a review of complaint files under a non-product capa and identified that a report was required to be filed as a result of the subject event.Distributed devices are not affected by the capa.Additionally, no new risks were identified, and no product-related actions are required as a result of this report submission.Furthermore, product performance trending is not impacted by this report, and no new mitigation actions are required as a result of this report.
|