No complaint was received with return of the device.Failure event observed during analysis.Final analysis revealed silicone material inside the aring, atip, lvtip, and rv/svc set screw hex cavities.This material prevented full insertion of the torque driver in the hex cavity and resulted in difficulty tightening the set screw.Abbott conducted a review of complaint files under a non-product capa and identified that a report was required to be filed as a result of the subject event.Distributed devices are not affected by the capa.Additionally, no new risks were identified, and no product-related actions are required as a result of this report submission.Furthermore, product performance trending is not impacted by this report, and no new mitigation actions are required as a result of this report.
|