| Model Number |
3667 |
| Medical Device Problem Code |
Application Program Freezes or Fails to Launch (4031)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
10/31/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
|
| |
|
Event or Problem Description
|
|
This report is related to a belgium patient.It was reported the patient's ipg became inoperable after coming in contact with bipolar electrocautery.As a result, the patient may undergo surgical intervention.
|
| |
|
Event or Problem Description
|
|
Follow-up revealed the patient's ipg was explanted and replaced (with a different model) to address the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: date of explant listed on follow-up report number 1 was determined to be incorrect upon further review.
|
| |
|