Catalog Number 71674071
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/01/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery the guide bolt wrench head snapped off inside the guide bolt.All pieces accounted for.Delay of over 2 hours reported.No injury reported.Procedure finished with competitor device.
 
Additional Manufacturer Narrative
The associated guide bolt and guide bolt wrench were returned and evaluated.A visual inspection found that the hex head of the wrench has sheared off; the broken piece was returned with the device.The hex feature can break if the torque applied to the driver exceeds the material strength.The device was manufactured in 2007.The guide bolt was stuck inside the drill guide and could not be removed.The date of manufacture is unknown for this part.Our investigation did not determine a specific cause of the stated failure.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
7/16 GUIDE BOLT
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8104357
Report Number1020279-2018-02595
Device Sequence Number1190955
Product Code HSB
UDI-Device Identifier03596010560643
UDI-Public03596010560643
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71674071
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/02/2018
Supplement Date Received by Manufacturer11/02/2018
Initial Report FDA Received Date11/26/2018
Supplement Report FDA Received Date01/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-