| Catalog Number |
71674071 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during surgery the guide bolt wrench head snapped off inside the guide bolt.All pieces accounted for.Delay of over 2 hours reported.No injury reported.Procedure finished with competitor device.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated guide bolt and guide bolt wrench were returned and evaluated.A visual inspection found that the hex head of the wrench has sheared off; the broken piece was returned with the device.The hex feature can break if the torque applied to the driver exceeds the material strength.The device was manufactured in 2007.The guide bolt was stuck inside the drill guide and could not be removed.The date of manufacture is unknown for this part.Our investigation did not determine a specific cause of the stated failure.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
|
| |
|
Search Alerts/Recalls
|