• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523LNAB
Medical Device Problem Codes No Display/Image (1183); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 03/12/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was received with blank display, problem isolated to lcd board.Unable to perform operating currents, self-test, unexpected restart error, rewind, basic occlusion, occlusion, prime/compromised force sensor system, excessive no delivery and displacement test due to blank display.The device had broken battery tube threads.
 
Event or Problem Description
Customer reported via phone call that they experienced high blood glucose level.Customer's blood glucose level was at 392 mg/dl the time of incident.Customer treated high blood glucose level with manual injection.Customer did not report any symptoms related to the high blood glucose level.Customer reported that the insulin pump had blank display.Customer stated that the battery compartment was cracked.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAB
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8104392
Report Number2032227-2018-66771
Device Sequence Number11267104
Product Code OYC
UDI-Device Identifier00643169151277
UDI-Public(01)00643169151277
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAB
Device Catalogue NumberMMT-523LNAB
Device Lot NumberA2523LNABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2018
Initial Report FDA Received Date11/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age13 YR
Patient Weight45
-
-