No complaint was received with return of the device.Failure event observed during analysis.The device was returned in hardware backup mode due to a por reset.The device image was retrieved but was corrupted.Upon interrogation, the battery voltage dropped suddenly and telemetry was lost.The cause of the backup mode could not be determined.Abbott conducted a review of complaint files under a non-product capa and identified that a report was required to be filed as a result of the subject event.Distributed devices are not affected by the capa.Additionally, no new risks were identified, and no product-related actions are required as a result of this report submission.Furthermore, product performance trending is not impacted by this report, and no new mitigation actions are required as a result of this report.
|