| Model Number |
LF1744 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the hand instrument was connected to the generator and the sealing could be performed without problem.However, in the middle of the operation, the fragments like insulation part fell into the patient's cavity and all the fragments were collected by forceps.To deal with the event, a new device was opened and used without problem.No patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information:evaluation summary: one device was received for evaluation.The returned product did not meet specification as received.Visual inspection found the bottom jaw over mold was damaged and missing plastic.The disengaged plastic was not returned.The reported condition was confirmed.The investigation found the bottom jaw over mold was damaged and missing plastic near the hinge of the device.The damage to the jaw's over mold likely occurred during the insertion or extraction of the device jaw from a trocar instrument.The investigation identified the root cause of the reported event to be user error.The instruction for use states, carefully insert and withdraw the instrument through the cannula to avoid damage to the device and or injury to the patient.Close jaws using device handle before insertion/extraction in the trocar.Corrective actions have been implemented to mitigate this issue.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|