• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO SPRINTER LEGEND OTW; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO SPRINTER LEGEND OTW; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Catalog Number SPL12510WL
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Codes Occlusion (1984); No Consequences Or Impact To Patient (2199); Patient Problem/Medical Problem (2688); Vascular Dissection (3160)
Date of Event 10/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An attempt was made to use one sprinter legend otw ptca balloon catheter to treat a lesion in the lad.It was reported that a balloon rupture occurred.
 
Additional Manufacturer Narrative
The target lesion was moderately calcified.The balloon was inspected and negative prep was performed prior to use with no issue noted.There was some degree of resistance in advancing the balloon catheter across the lesion.Excessive force was not used.The balloon rupture occurred during first inflation at nominal pressure.The device was not moved or repositioned in the lesion prior to the rupture.Blood was noted in the indeflator when device when placed on negative pressure.Dissection of the lad was noted after the inflation.Loss of flow was noted and the decision was made to place an iabp and send the patient to the or for bypass, given that the lesion could not be dilated.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINTER LEGEND OTW
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
MDR Report Key8105592
Report Number9612164-2018-03366
Device Sequence Number16634588
Product Code LOX
UDI-Device Identifier00643169854949
UDI-Public00643169854949
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K103095
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/03/2021
Device Catalogue NumberSPL12510WL
Device Lot Number215588453
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/13/2018
Supplement Date Received by Manufacturer11/29/2018
01/29/2019
Initial Report FDA Received Date11/27/2018
Supplement Report FDA Received Date12/17/2018
01/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient Weight56
-
-