• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted; give date: n/a (not applicable).The intraocular lens was not implanted.If explanted; give date: n/a (not applicable).The intraocular lens was not implanted and therefore not explanted.(b)(6).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the pcb00 +17.0 diopter tecnis monofocal iol (intraocular lens) did not come out of the inserter after the second push of the plunger.There was patient contact with the cartridge tip and the lens.Another pcb00 lens was implanted.It was also noted that there was no patient injury and no incision enlargement.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes, returned to manufacturer on: 01/14/2019.Device returned to manufacturer? yes.Device evaluation: the preloaded delivery system, model pcb00, was returned at the manufacturing site for evaluation.Visual inspection under the microscope showed that the intraocular lens (iol) was stuck in the cartridge.The plunger and pushrod were observed in advanced position.Residue of viscoelastic material was observed on the cartridge.No damage was observed on the cartridge.No assembly error and/or defect were observed in the preloaded device related to manufacturing process.The condition of the returned sample was consistent with a product that was handled and prepared for a surgical process.The customer's reported complaint was verified.However, there is no evidence to suggest that the complaint sample has been affected by the manufacturing process.No product deficiency was identified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key8105616
Report Number2648035-2018-01518
Device Sequence Number16650726
Product Code HQL
UDI-Device Identifier05050474558205
UDI-Public(01)05050474558205(17)210312
Combination Product (Y/N)Y
Initial Reporter CountryNL
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/05/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/12/2021
Device Model NumberPCB00
Device Catalogue NumberPCB0000170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2019
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/07/2018
Supplement Date Received by Manufacturer01/22/2019
10/25/2020
Initial Report FDA Received Date11/27/2018
Supplement Report FDA Received Date01/25/2019
11/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-