CONMED CORPORATION 4.5MM POPLOK SUTURE ANCHOR; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Model Number |
CKP-4500 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complaint is confirmed.Visual examination found the anchor implant broken and still attached to the shaft.The shaft driver was damaged (bent) due to excessive force and the device trigger was not pressed in.Visual evaluation was performed per assembly print.Excessive twisting forces may have attributed to this failure.The manufacturing documents from the device history record (dhr) have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution were found to have met all specifications prior to shipment.A dhr review found no abnormalities that would contribute to this issue.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 31 complaints, regarding 32 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised the following: warnings: surgeon must choose proper implant size based on specific procedure and patient history.Precautions: ensure proper selection of bone punch according to anchor size selection.Avoid lateral loading while inserting the poplok knotless suture anchor.Maintain proper alignment during insertion of the anchor and disengagement of the driver.Proper orientation and alignment of instruments is important during implantation of the poplok knotless suture anchor to minimize possible breakage of the anchor.Breakage of the poplok knotless suture anchor is possible if: the bone punch is not inserted to the proper depth; the poplok knotless suture anchor is not properly aligned with the pilot hole; the poplok knotless suture anchor inserter is used for prying.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
The customer reported an issue with the 4.5mm poplok suture anchor, item ckp-4500, lot # 948639 that occurred during a rotator cuff repair surgery involving a (b)(6) yr old male on (b)(6) 2018.It was reported that after making a pilot hole using pkl-45m, the surgeon passed sutures through the poplok anchor using suture loading tab and advanced the anchor head into the pilot hole using a mallet.After that, the surgeon noticed that the anchor was broken inside of the patient's body due to the hard bone.The broken piece was removed using a grasper.There was no report of impact or injury to the patient.The surgery was completed successfully with a 5-minute delay using an alternative foot print device.This report is raised on the basis of malfunction with potential for injury upon reoccurrence.
|
| |
|
Search Alerts/Recalls
|
|
|