| Medical Device Problem Codes |
Structural Problem (2506); Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a temperature (temp - physical damage) issue.It was reported that the battery compartment was damaged.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because there is the potential for the user to experience an injury to the skin due to increased pump temperature.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 04-oct-2019 with the following findings: during investigation, there was no activity related to pi observed in black box or download history.There was no voltage drops in black box.No overheating duplicated during testing.Pump electrical current draws found within specification.The battery compartment is cracked.Corrosion in battery compartment.Pump leaks at battery compartment.Pump opened; no further evidence of moisture found.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
| |
|