Pc-(b)(4).Batch # r92j4n.Device analysis: the analysis results found that the msm20 device was returned with no damage in the external components.In addition, the tyvek was returned along with the instrument.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled and it fed and formed the remaining 5 clips as intended.As the device was found to be fully functional, it could not be determined what may have caused the reported incident.No conclusion could be reached as to what may have caused the reported incident.The reported complaint could not be confirmed.The lot history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the packaging process.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.Additional information requested and received: did the clip drop from the jaws of the device? yes.Did the clips fall off the vessel after being placed? yes.
|