| Model Number |
N/A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
05/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Exemption number e2016018 b.Braun medical, inc.(importer) is submitting this report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as b.Braun melsungen ag internal report # (b)(4).The device is currently on shipping from brazil to bbm in germany for investigation.A follow-up report will be provided after the inspection results are available.Device history record (dhr): /reviewed the complaint batch and no abnormality found during in-process and at final control inspection.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility (translation of user facility information by bbm sales organization in brazil): broken/ needle.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number e2016018.(b)(4).This report has been identified as b.Braun melsungen ag internal report # (b)(4).We received one used spinocan 27gx3 1/2" (0.42x88 mm) in open packaging.The received sample was taken to a visual inspection.The used spinocan cannula is broken off approx.43 mm away from the cannula hub and the part with the cannula tip is bent.The structure of the break of the raw cannula shows that the cannula was repeatedly bent before the break.In addition, this shape resembled the shape of a broken cannula deliberately bent.Anomalies in which cannula folds and bends do not occur in the manufacturing process and transportation.In addition, the outside diameter of the spinocan cannula was measured according to drawing.Nominal-value: 0.42 +0.01/-0 mm.Actual-value: used sample = 0.42 mm.The measured value (outside diameters) of the spinocan cannula is in accordance with our specification.With regard to the used sample we assume of a problem during application and consider the complaint not confirmed.
|
| |
|
Search Alerts/Recalls
|