| Model Number |
SD300BA |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
09/15/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product was not returned therefore no analysis could be performed.Root cause could not be determined.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the bronchoscopy adaptor required excessive force to remove from the bronchoscope, which caused damage to the bronchoscope, loosening the instrument channel, allowing biomatter to accumulate under the device's o-ring, which cannot be cleaned, disinfected and/or removed during reprocessing.It allegedly contributed to a patient infection.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Patient was seen in (b)(6) 2016, diagnosed with dermatomyositis and started on prednisone, methotrexate and hydrochloroquine.Patient was then seen later for multiobar pna, pseudomonas.Patient received a chest x-ray and oxygen.Patient was treated with zosyn in the hospital and discharged with levaquin for 14 days.
|
| |
|