The pipeline flex was returned for evaluation without the catheter.The distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.The distal hypotube and ptfe shrink tubing were found to be intact with no signs of elongation.The pipeline flex braid was returned detached from the pushwire; therefore, the distal and proximal ends of the braid were unable to be identified.The pipeline flex braid appeared to be fully opened and moderately frayed at both ends.No bends were observed on the pushwire.No damages were found with the tip coil, distal marker, re-sheathing marker, re-sheathing pad or with the proximal bumper.No other anomalies were observed.Based on the analysis findings and the customer's photos, the customer reports of "failure to open at the distal end" could not be confirmed.The event cause could not be determined.The pipeline flex braid found to be fully opened with moderately frayed at both ends.In addition, no damages were found with the pushwire.Possible contributing factor of "failure to open" includes patient tortuous anatomy.The damage to the ends of the pipeline flex braid is likely the results of the physician re-sheathing the device more than recommended two times.Per our instructions for use (ifu): ¿begin to deliver the device using a combination of unsheathing the pipeline flex embolization device and pushing delivery wire simultaneously.After the distal end of the pipeline flex embolization device has successfully expanded, deploy the remainder of pipeline flex embolization device by pushing the delivery wire and/or unsheathing the pipeline flex embolization device.Re-sheathing and/or manipulation of the micro catheter, by locking down the delivery wire and moving both as a system, may facilitate expansion of the pipeline flex embolization device.Carefully inspect the deployed pipeline flex embolization device under fluoroscopy to confirm that it is completely apposed to the vessel wall and not kinked.If the device is not fully apposed or is kinked, consider using a balloon catheter, micro catheter, or guide-wire to fully open it.Do not use in patients in whom the angiography demonstrates the anatomy is not appropriate for endovascular treatment, due to conditions such as severe intracranial vessel tortuosity or stenosis.¿ no evidence was found to suggest that the devices failed to meet specifications; therefore, manufacturing has been ruled out as a potential cause.All products are 100% inspected for damages and irregularities during manufacture.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|