• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LUND MONITORS PRISMAFLEX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LUND MONITORS PRISMAFLEX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 115269
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Injury (2348)
Event Date 11/04/2018
Event Type  Injury  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a prismaflex machine removed 224ml of fluid from a patient within an hour, while the patient was performing continuous renal replacement therapy.The fluid removal setting for therapy was set to zero.The nurse also reported flow problem alarms during the same hour.The patient was treated with unspecified vasopressors due to the unintended fluid removed and continued therapy for over 60 hours with no additional issues noted.The patient outcome was not reported.No additional information is available.
 
Manufacturer Narrative
The device was not returned for evaluation.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - LUND MONITORS
lund skane lan
MDR Report Key8107820
MDR Text Key128471425
Report Number9616026-2018-00025
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
PMA/PMN Number
K131516
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/27/2018,01/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/27/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number115269
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA11/27/2018
Distributor Facility Aware Date11/05/2018
Event Location Hospital
Date Report to Manufacturer11/27/2018
Date Manufacturer Received01/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PRISMAFLEX M100 SET; PRISMAFLEX M100 SET
Patient Outcome(s) Required Intervention;
-
-