COVIDIEN MFG SOLUTIONS S.A. PALINDROME; CATHETER, HEMODIALYSIS, IMPLANTED Back to Search Results
Model Number 8888145042P
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during treatment, the flow in the dialysis line couldn't be taken.The a side and v side were reversely connected, but the bloody removal failed.The blood removal failure occurred in less than a week after the catheter was placed in the left femoral of the patient.It was stated that the catheter could be used for blood return, but it couldn't be used for blood removal, and the a side line was taken by dialysis needle.Flushing was performed prior to use and the device was flushed without problems.The catheter was replaced.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection and functional testing confirmed there were no abnormalities that would have caused or contributed to the reported condition.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during treatment, the flow in the dialysis line couldn't be taken.The a side and v side were reversely connected, but the blood removal failed.The blood removal failure occurred in less than a week after the catheter was placed at the left femoral of the patient.It was stated that the catheter could be used for blood return, but it couldn't be used for blood removal, and the a side line was taken by dialysis needle.Flushing was performed prior to use and the device was flushed without problems.The catheter was replaced.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALINDROME
Common Device Name
CATHETER, HEMODIALYSIS, IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
MDR Report Key8108057
Report Number3009211636-2018-00374
Device Sequence Number1827620
Product Code MSD
UDI-Device Identifier20884521158051
UDI-Public20884521158051
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K123196
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8888145042P
Device Catalogue Number8888145042P
Device Lot Number8888145042P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2018
Initial Date Received by Manufacturer 11/03/2018
Supplement Date Received by Manufacturer02/15/2019
Initial Report FDA Received Date11/27/2018
Supplement Report FDA Received Date03/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-