Model Number UNKNOWN-S |
Device Problems
Mechanics Altered (2984); Insufficient Information (3190)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/01/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that prior to implant, the patient was in a vegetative state following a car accident that caused a brain hemorrhage in (b)(6) 2015.In (b)(6) 2016, the patient was implanted due to hydrocephalus and being in a coma.Towards the end of (b)(6) 2018, the patient underwent an mri and a routine pressure adjustment was needed after the mri.The physician was to arrange to have the pressure adjusted and the patient was still in a vegetative state.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received reported the pressure setting of the valve did change following the mri.The valve was readjusted according the doctor¿s requirement and the patient was in good condition.
|
|