Catalog Number SXPP1A200
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pc-(b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Investigation summary: it was reported performance pull off suture needle pre-op.An empty opened folder, an empty labeled winding former, a dispensed suture and a detached needle were returned for analysis.During the visual inspection of the needle the swage and attachment area were as expected.The barrel hole of the needle was examined under 20x magnification and no suture remnant was noted.The suture was examined, and the insertion did not meet the requirement.  the manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Per the sample condition, the assignable cause of the performance pull off suture needle pre-op is a short insertion.
 
Event or Problem Description
It was reported that a patient underwent a total knee arthroplasty on (b)(6) 2018 and a barbed suture was used.During the procedure, the suture was detached from the needle when holding the needle with a needle holder.It was not used for the patient.There were no adverse patient consequences reported.Upon evaluation, the insertion did not meet the requirement.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFX SYM PDS+ VIO 18IN 0 S/A OS-6
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key8108203
Report Number2210968-2018-77360
Device Sequence Number19350383
Product Code NEW
UDI-Device Identifier10705031227668
UDI-Public10705031227668
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K113004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Catalogue NumberSXPP1A200
Device Lot NumberLMK296
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/30/2018
Initial Report FDA Received Date11/27/2018