• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS 8.0FR W/-4 VITACUFF PERC; CHRONIC CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS 8.0FR W/-4 VITACUFF PERC; CHRONIC CATHETER Back to Search Results
Model Number 7711804
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or no medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Event or Problem Description
It was reported the cuff allegedly detached but remained on the chronic catheter during insertion.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one 8fr groshong catheter was returned for evaluation.Visual inspection identified a loose tissue cuff which could freely move up and down the returned catheter.The adhesive residue was noted at the correct location on the 24 cm depth marker.Therefore, the investigation is confirmed for the alleged defective tissue cuff.The definitive root cause could not be determined based upon the available information.It is unknown if patient and/or procedural issues contributed to the reported event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(expiration date: 07/2023).
 
Event or Problem Description
It was reported the cuff allegedly detached but remained on the chronic catheter during insertion.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8.0FR W/-4 VITACUFF PERC
Common Device Name
CHRONIC CATHETER
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8108254
Report Number3006260740-2018-03412
Device Sequence Number1997317
Product Code LJS
UDI-Device Identifier00801741036767
UDI-Public(01)00801741036767
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K831386
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7711804
Device Catalogue Number7711804
Device Lot NumberRECU2174
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2018
Initial Date Received by Manufacturer 11/01/2018
Supplement Date Received by Manufacturer01/12/2019
Initial Report FDA Received Date11/27/2018
Supplement Report FDA Received Date02/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-