| Catalog Number |
6260-9-144 |
| Medical Device Problem Codes |
Degraded (1153); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Edema (1820); Inflammation (1932); Injury (2348); Reaction (2414)
|
| Date of Event |
10/30/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Lot specific voluntary recall v40 - recall - 2249697-05/07/2018-003r was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.Device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that patient's right hip was revised due to disassociation.Intra-operatively, pseudo tumor was reported with an excessive amount of black tissue observed.The trunnion was found to be worn down and the liner was worn through in at least 2 places.The shell was well-fixed and not revised.Patient was revised to a restoration modular stem construct with a 28 +8 metal head.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding disassociation and altr involving a metal head mated with an accolade stem was reported.A review by a clinician of the provided x-rays confirmed disassociation.Review of the provided intra-operative photos showed metallosis but histopathology reports would be required in order to confirm altr.Method & results: product evaluation and results: the device was not returned but photos of the explant were provided.There is some biological material on the outer surface of the shell and a small amount of black debris inside the taper of the head.Medical records received and evaluation: provided x-rays, explant photos and intra operative photo were reviewed by a clinician who gave the following comment: provided x-ray showed dissociated components, photographs showed metallosis and trunnion deformities, need operative reports, histopathology reports, serial x-rays, office reports and examination of explanted components.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there were no other events for the lot provided.Conclusions: lot specific voluntary recall v40 - recall - 2249697-05/07/2018-003r was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before (b)(6) 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
|
| |
|
Event or Problem Description
|
|
It was reported that patient's right hip was revised due to disassociation.Intra-operatively, pseudo tumor was reported with an excessive amount of black tissue observed.The trunnion was found to be worn down and the liner was worn through in at least 2 places.The shell was well-fixed and not revised.Patient was revised to a restoration modular stem construct with a 28 +8 metal head.
|
| |
|