| Model Number |
760 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the service engineer evaluated the ventilator and replaced the proportional solenoid (psol) printed circuit board (pcb).After serv icing, the ventilator passed all testing per manufacturer's specification and was returned to the customer.The psol pcb was visually inspected with no anomalies observed.The component was connected to a test ventilator.The fault was isolated to component u11.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the 760 ventilator had failed sst(short self-test).The ventilator was not in use on a patient at the time of the reported event. .
|
| |
|