| Catalog Number |
DSF2633 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
10/24/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the manufacturing records is being requested.According to the gore® dryseal flex sheath instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to: blood loss and vascular trauma.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2018, a patient underwent endovascular treatment in the thoracic aorta, zone 0, with gore® tag® thoracic branch endoprostheses and a conformable gore® tag® thoracic endoprosthesis as distal extension to aortic component.According to the report all devices were positioned and implanted without issue.A gore® dryseal flex sheath was used for access.During the procedure an iliac artery rupture occurred.Estimated blood loss was 2500ml.Blood replacement products were given and a right femoral artery repair was performed.The patient tolerated the procedure.Post-operatively the patient had low blood pressure, respiratory insufficiency and atrial fibrillation, which were all medically managed.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the manufacturing records for the device verified the lot met all pre-release specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
According to the gore® dryseal flex sheath instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to: blood loss and vascular trauma.Additionally, the sizing for a 26fr sheath requires a minimum access vessel diameter of 8.7mm.This patient's access vessel diameter was 6.8-8.7mm.
|
| |
|