MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Catalog Number 55840007545
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was pre-operatively diagnosed with pseudoarthrosis and underwent screw removal at s1, anterior lumbar interbody fusion (alif) post fusion at t5-ilium.Post-op, screw was found to be broken.The patient did not achieve fusion.Revision surgery was performed for non-fusion.No fragment remains in the patient.Patient complications reported to be unknown.
 
Additional Manufacturer Narrative
Product analysis: visual inspection confirmed the head of the mas screw has been broken off and not returned for analysis.Microscopic inspection of the fracture surface of the remaining bone screw revealed multiple types of damage that is not analyzable.The threads near the top of the fracture are damaged/bent.The above observations are consistent with the damages done from the screw removal process.Root cause of original failure unknown.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key8108913
Report Number1030489-2018-01564
Device Sequence Number16619429
Product Code KWP
UDI-Device Identifier00643169004344
UDI-Public00643169004344
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K113174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number55840007545
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2019
Initial Date Received by Manufacturer 10/29/2018
Supplement Date Received by Manufacturer01/18/2019
Initial Report FDA Received Date11/27/2018
Supplement Report FDA Received Date01/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age71 YR
-
-