| Model Number |
18785 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, upon inflation, the device's cuff did not stay inflated.It was reported that the test was redone prior to removal of the tube from packaging for further review.Customer reported that there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one sample device was received for evaluation.The reported event in cuff was confirmed.The visual inspection was performed, and it was observed the cuff presents a cut.An inflation/deflation test was performed and it was observed the cuff deflated immediately.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|