JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZMB00
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Unintended Movement (3026)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Explant date: if explanted, give date: not applicable, as the lens remains implanted.  (b)(4).It was indicated that the device is not returning for evaluation as it remains implanted; therefore a failure analysis of the complaint device cannot be completed.A review of the device/lot history record, labeling, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson and johnson surgical vision has been submitted.
 
Event or Problem Description
It was reported that the lens was implanted in patient¿s right eye that had capsular tear.Lens dislocated post operatively and was noticed on day 54 post implantation.The lens was planned to be explanted in secondary surgery because of patient experiencing cloudy vision, double vision and decreased visual acuity.Reportedly the lens was scheduled to be explanted on (b)(6) 2018.Additional information was received, and it was learnt that surgery was aborted due to unstable capsule.One 10-0 suture was used on the primary incision which was not enlarged.Patient was doing fine.The lens is still implanted.Visual acuity pre-op - 20/400 bcva (best corrective visual acuity).Visual acuity post-op - 20/100, pinhole 20/60.No further information provided.
 
Additional Manufacturer Narrative
The product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.The manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the review.The product was manufactured and released according to specifications.A search revealed that no other complaints were received for this production order.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8109533
Report Number3011852734-2018-00221
Device Sequence Number16709588
Product Code MFK
UDI-Device Identifier05050474535626
UDI-Public(01)05050474535626(17)220306
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/05/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/06/2022
Device Model NumberZMB00
Device Catalogue NumberZMB00U0200
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/31/2018
Supplement Date Received by Manufacturer12/17/2018
10/25/2020
Initial Report FDA Received Date11/28/2018
Supplement Report FDA Received Date12/28/2018
11/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Date Report Sent to FDA01/01/2005
-
-