| Catalog Number |
873-004 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was pre-operatively diagnosed with odontoid process fractures and underwent anterior screw fixation surgery.During surgery, the guide wire broke.The product came in contact with the patient.There was a delay of less than 60 mins in procedure time as a result of the event.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: visual and optical inspection confirmed about an inch of the guidewire has been broken off.Optical inspection of the fracture surface confirmed a brittle surface with multiple ridges and some circular material flow.This type of damage is consistent with torsional overload causing the tip to break off.If information is provided in the future, a supplemental report will be issued.
|
| |
|