| Catalog Number |
48811228 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Pain (1994); Injury (2348)
|
| Date of Event |
11/07/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The bottom screws backed out and the locking mechanism was pushed out and snapped on the anterior cervical plate.The original surgery was approximately 2 years ago and the patient called the doctor complaining of neck pain.Revision surgery occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: x-ray image review, nc/capa history review, labelling review, risk assessment.Result: the reported event was confirmed via received x-ray image.Device was not returned, therefore, an evaluation could not be performed.No lot # was provided, so a manufacturing record review could not be performed.Conclusion: the root cause of the reported event was determined to be backing out of bottom screws of the plate which resulted due to continuous load on the device for 2 years.
|
| |
|
Event or Problem Description
|
|
The bottom screws backed out and the locking mechanism was pushed out and snapped on the anterior cervical plate.The original surgery was approximately 2 years ago and the patient called the doctor complaining of neck pain.Revision surgery occurred.
|
| |
|