| Catalog Number |
48770820 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tip of a quick turn screw driver broke inside of screw head during screw insertion.A surgical delay of 30 minute was reported.Broken fragments were retrieved and surgery was completed successfully.No other adverse consequence to the patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection, product history review, complaint history review, nc/capa history review, labelling review, risk assessment result: the reported event was confirmed upon visual inspection.Manufacturing files were reviewed and no issues were found.The device was found to be over 8 years old.Additionally, it was reported that the instrument had been used 2-3 times per month over the last two years a conclusion: the probable root cause of the tip fracture is the age of the instrument (found to be over 8 years old).
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of a quick turn screw driver broke inside of screw head during screw insertion.A surgical delay of 30 minute was reported.Broken fragments were retrieved and surgery was completed successfully.No other adverse consequence to the patient was reported.
|
| |
|