| Model Number |
MMT-723NAH |
| Medical Device Problem Codes |
Moisture or Humidity Problem (2986); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pump received with blank display due to moisture damage at electronic assembly.Also, moisture damage motor during visual inspection.Unable to perform operating currents, self test, rewind test and displacement test due to blank display.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump was damaged.Customer blood glucose level was unknown.Customer also stated that the top and bottom of insulin pump was cracked and insulin pump was also exposed to moisture.The device will be returned for analysis.
|
| |
|