Model Number BURR, OVAL, 12 FLUTE 5.5MM X 13CM
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Codes Burn(s) (1757); Burn, Thermal (2530)
Date of Event 11/09/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was requested/is expected but has not returned for evaluation.The cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that the oval burr, ar-8550obt, was being used in a case and shortly after it began spinning, the burr became hot and burned the patient's skin.Additional information provided (b)(6) 2018: this event occurred while the surgeon was performing a subacromial decompression during a shoulder arthroscopy procedure.The patient was burned on their shoulder; however, no photos were captured.Additional information provided (b)(6) 2018: a photo was taken at the patient's follow-up appointment.
 
Additional Manufacturer Narrative
The complaint is confirmed.After visually evaluating the picture of a patient attached to the complaint, it is concluded that the device caused damage/burn in the tissue of the patient.The device was not returned for evaluation.The cause is attributed to a supplier process and two root causes were identified: material and measurement.Material: material is considered to be a root cause as components that did not meet specification contributed to the event.Because the heat shrink recovered outer diameter and the outer tube inner diameter were oversized and undersized respectively, proper clearance did not exist between the burr components which cause the generation of heat.Measurement: measurement is considered to be a root cause due to the inadequate ""spins freely"" test which allowed nonconforming material to be accepted into inventory.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURR, OVAL, 12 FLUTE 5.5MM X 13CM
Common Device Name
BURR, SURGICAL, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key8110257
Report Number1220246-2018-00769
Device Sequence Number16735245
Product Code GFF
UDI-Device Identifier00888867044050
UDI-Public00888867044050
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model NumberBURR, OVAL, 12 FLUTE 5.5MM X 13CM
Device Catalogue NumberAR-8550OBT
Device Lot Number10211276
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2018
Supplement Date Received by Manufacturer11/09/2018
Initial Report FDA Received Date11/28/2018
Supplement Report FDA Received Date06/07/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/14/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;