Model Number FSC330-F7523-R
Medical Device Problem Code Image Display Error/Artifact (1304)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Although expected, the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a fuse c38 colonoscope - r was used in the colon during a colonoscopy performed on (b)(6) 2018.According to the complainant, during the procedure, the images flickers when angulating the scope.The anatomy of the patient is not visible in the center image when the image flickers.Another fuse c38 colonoscope - r was used to complete the procedure.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
H6: device code 1304 captures the reportable event of center image lost.H10: a fuse c38 colonoscope - r was received for analysis.A functional evaluation was performed and found an internal wiring failure of the charged couple device (ccd) resulting to loss of all images when angulated.The reported issue was confirmed.The cause of the reported failure is due to ccd failure.Therefore, the assigned investigation conclusion code for this complaint is designated as cause traced to component failure.The repair history was reviewed and nothing was found to indicate a possible service-related cause for the complaint.
 
Event or Problem Description
It was reported to boston scientific corporation that a fuse c38 colonoscope - r was used in the colon during a colonoscopy performed on (b)(6) 2018.According to the complainant, during the procedure, the images flickers when angulating the scope.The anatomy of the patient is not visible in the center image when the image flickers.Another fuse c38 colonoscope - r was used to complete the procedure.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSE
Common Device Name
ENDOSCOPE, ACCESSORIES, IMAGE POST-PROCESSING FOR COLOR ENHANCEMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8110306
Report Number3005099803-2018-61607
Device Sequence Number2364088
Product Code PEA
UDI-Device Identifier00840253114372
UDI-Public00840253114372
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K160275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFSC330-F7523-R
Device Catalogue Number60093
Device Lot Number2300226
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2018
Initial Date Received by Manufacturer 11/05/2018
Supplement Date Received by Manufacturer01/10/2019
Initial Report FDA Received Date11/28/2018
Supplement Report FDA Received Date01/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-