Model Number SCD48
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during enterotomy in a gastric bypass, the device made an error sound and the tip of the blade broke off into the patients cavity.They were able to retrieve the blade.They finished the procedure using a different kind of device.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: evaluation summary: one device was returned for evaluation.The returned product did not meet specification as received.Visual inspection of the disposable device revealed that the tip of the waveguide was broken off.The broken piece was returned.Witness marks indicate that the device had been used.The reported condition was confirmed.The device was returned with a fractured waveguide.The device was not functional.Investigation personnel concluded that the titanium waveguide cracked from continued use after becoming damaged, and may have come in contact with any of the following; hemostats, clips, staples, retractors, etc.During use.The broken piece was returned.The investigation identified the cause of the reported event to be user error.The device instructions for use currently contain a warning against contact between the dissector tip and metal objects (hemostats, clips, staples, retractors, etc.) during activation.Contact with metal objects during use will cause the active blade to crack and may eventually break off.This issue is specific to ultrasonic dissectors.The instructions for use (ifu) advises device users to visually inspect all system components for breaks, cracks, nicks, or other damages prior to use.Ifu also states: do not use damaged components.Use of damaged components may result in injury to the patient or user.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONICISION
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key8110503
Report Number1717344-2018-01555
Device Sequence Number2261297
Product Code LFL
UDI-Device Identifier10884521195226
UDI-Public10884521195226
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K141371
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSCD48
Device Catalogue NumberSCD48
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2018
Initial Date Received by Manufacturer 11/05/2018
Supplement Date Received by Manufacturer11/30/2018
Initial Report FDA Received Date11/28/2018
Supplement Report FDA Received Date12/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-