• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 BIPHASIC LIFEPAK(R) 500 AUTOMATED EXTERNAL DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 BIPHASIC LIFEPAK(R) 500 AUTOMATED EXTERNAL DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Catalog Number 99401-000618
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control provided the customer with technical assistance and advised the customer that their device is no longer under warranty and there is no repair service for the device.Physio recommended that the customer replace the device and remove from service.The device has not been returned to physio-control.The cause of the reported issue could not be determined.Device not evaluated by manufacturer.
 
Event or Problem Description
The customer contacted physio-control to report that their device had logged an event code in the memory which is related to a potential issue of the device to operate properly.In this state, defibrillation therapy may be delayed or not available if needed.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPHASIC LIFEPAK(R) 500 AUTOMATED EXTERNAL DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key8110862
Report Number0003015876-2018-01998
Device Sequence Number10824174
Product Code MKJ
UDI-Device Identifier00721902549540
UDI-Public00721902549540
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K052057
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number99401-000618
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2018
Initial Report FDA Received Date11/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-