• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC EXP 4.5 TI SINGLE INNIE; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC EXP 4.5 TI SINGLE INNIE; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 186100001
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code Local Reaction (2035)
Date of Event 11/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Per reporter, mr (b)(6) was performing a distraction of growing rod procedure using the connector, he noticed that the set screw in the connector was not flush and upon removing the set screw and the connector he noticed some metallosis in the patients tissue.He replaced the connector and distracted the rods, with no issues.
 
Additional Manufacturer Narrative
(b)(4).Device was not returned for evaluation.A review of the device history record could not be performed as the lot number is unknown.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.Without the return of the device, we are unable to confirm the reported issue or identify the root cause.As no systemic trends were found, this complaint file will be closed with no further action required.Should more information and/or the sample be provided at a later time, this complaint will be reopened and device evaluated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXP 4.5 TI SINGLE INNIE
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
MDR Report Key8111012
Report Number1526439-2018-51094
Device Sequence Number2165819
Product Code NKB
UDI-Device Identifier10705034103846
UDI-Public(01)10705034103846
Combination Product (Y/N)N
PMA/510(K) Number
K160904
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number186100001
Device Catalogue Number186100001
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/06/2018
Supplement Date Received by Manufacturer12/11/2018
Initial Report FDA Received Date11/28/2018
Supplement Report FDA Received Date12/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-