| Model Number |
186100001 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
Local Reaction (2035)
|
| Date of Event |
11/06/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Per reporter, mr (b)(6) was performing a distraction of growing rod procedure using the connector, he noticed that the set screw in the connector was not flush and upon removing the set screw and the connector he noticed some metallosis in the patients tissue.He replaced the connector and distracted the rods, with no issues.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Device was not returned for evaluation.A review of the device history record could not be performed as the lot number is unknown.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.Without the return of the device, we are unable to confirm the reported issue or identify the root cause.As no systemic trends were found, this complaint file will be closed with no further action required.Should more information and/or the sample be provided at a later time, this complaint will be reopened and device evaluated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|