| Catalog Number |
4MAXC |
| Medical Device Problem Codes |
Break (1069); Device Damaged Prior to Use (2284)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the returned device was fractured approximately 104.0 cm from the hub.Conclusions: evaluation of the returned 4maxc confirmed a device that was fractured.If the 4maxc is forcefully retracted from the packaging at extreme angles or otherwise mishandled during preparation, damage such as a fracture may occur.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a thrombectomy procedure using a penumbra system 4max reperfusion catheter (4maxc).During the procedure, the physician placed a non-penumbra catheter in the vertebral artery (va).The physician then opened a 4maxc and noticed that it was fractured upon removal from the packaging; therefore, it was not used in the procedure.The procedure was completed using the same catheter, new 4maxc and a stent device.
|
| |
|