| Model Number |
M00558430 |
| Medical Device Problem Codes |
Material Puncture/Hole (1504); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient's exact age is unknown; however, it was reported that the patient was over the age of 18.(b)(6).(b)(4).Investigation results: a visual examination of the returned complaint device found that the exit marker was bunched up.A functional examination was performed, the balloon was inflated and a leak was found due to a pinhole located at the distal shoulder of the balloon.This failure is likely due to adverse factors and/or conditions related to the procedure during the use of the device which limited the performance of the device.Therefore, the most probable root cause is cause traced to component failure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2018.It was found that the balloon had a pinhole and the exit marker was bunched-up.Reportedly, the procedure was completed with another cre wireguided dilatation balloon.There were no patient complications reported as a result of this event.
|
| |
|