• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE WIREGUIDED; DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE WIREGUIDED; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00558430
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.(b)(4).Investigation results: a visual examination of the returned complaint device did not show visual defects.A functional examination was performed; the balloon was inflated, and it was found that the balloon was ruptured.This failure is likely due to encountered procedural factors and/or the manner as the device was handled and manipulated which could have affected its performance and its intended purpose, leading to the observed failure.Therefore, the most probable root cause is cause traced to component failure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilation balloon was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the balloon leaked when the physician tried to implant the stent.The procedure was completed with another cre wireguided dilation balloon.There were no patient complications reported as a result of this event.Investigation results revealed that the balloon burst; therefore this is now an mdr reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE WIREGUIDED
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8111252
Report Number3005099803-2018-61319
Device Sequence Number19414525
Product Code KNQ
UDI-Device Identifier08714729339366
UDI-Public08714729339366
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110833
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/04/2020
Device Model NumberM00558430
Device Catalogue Number45735
Device Lot Number0021089656
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2018
Initial Report FDA Received Date11/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-