• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION ALARIS PUMP MODULE ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS PUMP MODULE ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 24602-B007T
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product will be returned per customer.The customer complaint could not be confirmed because the product was discarded and not returned for failure investigation.The root cause of this failure was not identified.
 
Event or Problem Description
The customer reported that a leak occurred in the oncology unit during a pump infusion of a 500ml bag of normal saline containing etoposide 62.5mg, doxorubicin 12mg, and vincristine 0.35mg which was running at a rate of 46ml/hr.The bag was hung on (b)(6) 2018 at 0010 and the leak (from the center of the circular area near the end of the filter) was noted on (b)(6) 2018 at approximately 1115 after infusing for approximately 11 hours.There was no patient harm.
 
Additional Manufacturer Narrative
The photo provided by the customer confirmed the leak from the micron filter p/n 10012217 on the distal vent hole.The root cause of this failure was not identified.
 
Event or Problem Description
The customer reported that a leak occurred in the oncology unit during a pump infusion of a 500ml bag of normal saline containing etoposide 62.5mg, doxorubicin 12mg, and vincristine 0.35mg which was running at a rate of 46ml/hr.The bag was hung on (b)(6) 2018 at 0010 and the leak (from the center of the circular area near the end of the filter) was noted on (b)(6) 2018 at approximately 1115 after infusing for approximately 11 hours.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS PUMP MODULE ADMINISTRATION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key8111334
Report Number9616066-2018-02337
Device Sequence Number19265613
Product Code FPA
UDI-Device Identifier10885403240980
UDI-Public10885403240980
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K944320
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Model Number24602-B007T
Device Catalogue Number24602-B007T
Device Lot Number18046755
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/28/2018
Supplement Report FDA Received Date06/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
8100,8015, THERAPY DATE: (B)(6) 2018
Patient Age39 YR
Patient Weight68
-
-