• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT HI-TORQUE PILOT GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT HI-TORQUE PILOT GUIDE WIRE Back to Search Results
Catalog Number 1010480-HJ
Medical Device Problem Codes Product Quality Problem (1506); Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that a buildup of procedural contaminants (blood, contrast) on the guide wire resulted in the reported physical property issue (noted to feel sticky); thus resulting in the reported difficulty to position and the reported difficulty to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was performed to treat a de novo lesion in the mid right coronary artery with mild tortuosity, mild calcification and 80% stenosis.Following advancement of a hi-torque pilot 50 guide wire to the lesion, an unspecified balloon dilation catheter (bdc) was advanced for pre-dilatation.On removal of the bdc, resistance with the guide wire was met.An unspecified stent delivery system (sds) was attempted to be advanced over the same guide wire; however, resistance with the wire was met and the sds could not cross the lesion.Resistance was also met on removal of the sds, therefore, the guide wire and sds were removed together as a single unit.Following removal, the coating of the guide wire was noted to feel sticky.A new unspecified guide wire was used to successfully continue the procedure.While a delay in the procedure occurred, there were no reported adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE PILOT GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8111507
Report Number2024168-2018-09100
Device Sequence Number2364099
Product Code DQX
UDI-Device Identifier08717648073861
UDI-Public08717648073861
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K060449
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2020
Device Catalogue Number1010480-HJ
Device Lot Number8060671
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2018
Initial Report FDA Received Date11/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-